Fault Information:

Number
Name Data Field Condition Recovery
m9 Simulated m Fault
n/a User enters simulation mode
Reset simulated fault
m10
Over Range
Peak concentration Large concentration of gas has been detected. The Midas requires an independent confirmation that the gas hazard is gone.
Supply known clean air to the Midas and clear the fault.
m11
User Calibration Expired
Days until fault The user specified calibration interval has elapsed.
Perform zero and span calibrations.
Increase span calibration period.
(zero days = disabled)
m12
Cartridge Expires Soon
Days until fault Cartridge is old and will expire soon.
Replace the cartridge with a new cartridge.
m13
Flow Low
Flow (cc/min) Midas is no longer able to regulate the flow.
Check filters and pump.
m14
Interferent Present
A/D counts An interferent is degrading the ability of the Midas to detect gas. Check application.
m15
Temperature Near Limit
Temperature (C) Temperature is within 2 C of limit.
Check installation environment.
m16
Baseline Near Limit
A/D counts Sensor baseline has drifted.
Check for background gas concentration, temperature or humidity fluctuations.
Perform zero calibration.
Replace cartridge.
m17
Inhibit Timeout
n/a Transmitter has been in inhibit mode too long.
Resume monitoring or increase timeout value.
m18 Software Mismatch Software Version Coprocessor firmware or bootloader too old Update firmware
F39 Simulated F Fault
n/a User enters simulation mode
Reset simulated fault
F40
Sensor Overdosed
n/a Sensor has been exposed to high gas concentrations for long periods.
Replace cartridge.
F41
Baseline Fault
A/D counts Sensor baseline has drifted.
Check for background gas concentration, temperature or humidity fluctuations.
Perform zero calibration.
Replace cartridge.
F42
Calibration Expired
Code Too long since last calibration.
Replace or calibrate the cartridge.
F43
Cartridge Expired
Daughter code Cartridge is too old.
Replace cartridge.
F44
Cell Failure
Daughter code Cartridge has failed Reflex™ check.
Replace cartridge.
F45
Stabilization Timeout
A/D counts Cartridge has failed to stabilize.
If temperature or humidity shocks exist, precondition the cartridge.
Check for background gas concentration.
Replace cartridge.
F46
Cartridge Analog Failure
Code Over range or analog failure Ensure no dangerous gasses are present; replace cartridge.
F47
Cartridge Memory Invalid
Code Checksum error.
Replace cartridge.
F48
Cartridge Absent
n/a No communications.

Reseat cartridge.
Replace cartridge.

F49
Cartridge Type Changed
Type Cartridge part number changed from previous.
Verify cartridge is correct part number.
Replace if wrong part number.
Clear fault.
F80
Temperature Limits Exceeded
Temperature (C) Temperature is outside limits.
Check installation environment.
F81
Flow Fail
Flow (cc/min) Flow < 70% of nominal for 15 seconds
Check filters.
Check for kinked tubing.
Replace pump.
F82
Excessive Electrical Noise
n/a Internal electronics repeatedly noisy.

Check grounding of Midas chassis.
Check termination of cable shields.
Relocate the Midas further from noise sources.
Add ferrite inductors to cables.

F83
Pyro Comm Fail
n/a Pyrolyzer communication failure
Check electrical connection to pyrolyzer.
Reboot
Replace pyrolyzer.
F84 Misc. Transmitter Fault Daughter code Transmitter is defective. Service or replace Midas.
F85 FW Download Fail n/a Coprocessor FW Download Fail Retry downloading coprocessor FW.
F86 Coprocessor Fail n/a Coprocessor is damaged Contact Service.
Replace the coprocessor.
F87 Pyrolyzer Temperature Limits Exceeded Temperature (C) Pyrolyzer temperature exceeded Clean pyrolyzer cooling vents.
Contact Service.
F88 Pyrolyzer Heater Fail Resister (ohm) Pyrolyzer heater fail. Contact Service.
Replace pyrolyzer.
F89 Pyrolyzer Power Unreg Power (W) Target power not achieved within specified time span If 24 VDC power check,
21.6 < Vin < 26.4
Contact Service.
Replace pyrolyzer.
F90 Pyrolyzer Heater Mismatch Daughter code Incompatible with ver 6 firmware Pyrolyzer Contact Honeywell Analytics Service for further advice
F91 HTP Module Required Pyrolyzer Type HT Pyrolyzer is required for MIDAS-S-CFX or MIDAS-E-CFX cartridges. Change to HTP